This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ZYM B REAGENT (2015-12-23)

Starting date:
December 23, 2015
Posting date:
February 8, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57016

Affected Products

ZYM B REAGENT

Reason

Based on discrepant results (false negative results) reported from the field, bioMérieux initiated a complaint investigation. A loss of performance due to light exposure after 3 months of shelf life for API ZYM B, which lead to false negative results on some API strips using the ZYM B (API 20 Strep-20600, API Coryne-20900, API Staph-20500, API ZYM-25200, API NH-10400, API Listeria-10300).

Affected products

ZYM B REAGENT

Lot or serial number
  • 1003765610
  • 1003772410
  • 1003772470
  • 1003787970
  • 1003790030
  • 1004077210
  • 1004097610
  • 1004355410
  • 1004355730
Model or catalog number

70493

Companies
Manufacturer

Biomerieux Sa

Chemin De L'orme

Marcy-l'etoile

69280

FRANCE