Zoll ventilators with MRI compatibility
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Uni-Vent 731 EMV+ |
AY20A009538 |
731 EMV+ |
Eagle II Portable Ventilator |
VCCC100515 |
731 EAGLE II |
Z Vent |
AY19G008474 |
Z VENT |
Issue
Manufacturer is voluntarily recalling a limited number of operator's guides and quick reference guides related to Zoll ventilators with MRI compatibility. Through an internal review of product manuals, it was identified that documented patient safety ventilator MRI information was inadvertently omitted from the manual. The impacted manual only states to place the ventilator behind the 2000 gauss line. The updated language has been changed to: You must place the ventilator behind the 130 gauss field line (approximately 2 meters to the bore opening of a 3T MRI magnet). This update to the instruction ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy.
Recall Start Date: May 24, 2024
Additional information
Details
Zoll Medical Corporation
269 Mill Road, Chelmsford, Massachusetts, United States, 01824
Get notified
Receive emails about new and updated recall and safety alerts.