This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Zimmer Sterile Packaged Implants (2017-03-09)

Starting date:
March 9, 2017
Posting date:
March 24, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62780

Affected Products

A. M/Dn Intramedullar Femoral Recon Nail

B. Versys Hip Revision Femoral Stem Beaded

C. ITST Fixation Femoral Nail

D. Zmr Hip Xl Porous Stem Nitrided Bowed

E. Segmental System Fluted Stem Extension Bowed

F. Segmental System Fluted Stem Ext Straight

G. Segmental System Segment Male/Female Taper

H. Segmental System Variable Stiffness Stem Extension

Reason

Zimmer is initiating a field action for sterile-packaged implants in two different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the two configurations failed simulated distribution and shipping testing. The devices impacted are generally the heaviest outlier sizes within the respective product family.

Affected products

A. M/Dn Intramedullar Femoral Recon Nail

Lot or serial number

Tous les lots avant le 2027/03/01

Model or catalog number

00-2252-360-16

00-2252-380-16

00-2252-400-16

00-2252-420-16

00-2252-440-15

00-2252-440-16

00-2252-460-15

00-2252-460-16

00-2252-480-15

00-2252-480-16

00-2252-500-15

00-2252-500-16

00-2253-440-15

Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

UNITED STATES



B. Versys Hip Revision Femoral Stem Beaded

Lot or serial number

Before 2026/07 (All others)

Before 2027/03 (00784301508)

Model or catalog number

00784301508

00784301608

00784301708

00784301808

00784302008

Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

UNITED STATES



C. ITST Fixation Femoral Nail

Lot or serial number

All before 2027/03/01

Model or catalog number

00-2257-480-04

00-2257-480-14

00-2257-500-04

00-2257-500-14

Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

UNITED STATES



D. Zmr Hip Xl Porous Stem Nitrided Bowed

Lot or serial number

All before 2027/03/01

Model or catalog number

00-9921-240-33

00-9921-255-33

Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

UNITED STATES



E. Segmental System Fluted Stem Extension Bowed

Lot or serial number

Before 2026/07 (All others)

Before 2026/09 (00585205417)

Before 2027/03 (00585205415)

Model or catalog number

00-5852-054-15

00-5852-054-16

00-5852-054-17

00-5852-054-18

00-5852-054-19

Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

UNITED STATES



F. Segmental System Fluted Stem Ext Straight

Lot or serial number

All lots before 2027/03/01

Model or catalog number

00-5852-052-17

00-5852-052-18

00-5852-052-19

Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

UNITED STATES



G. Segmental System Segment Male/Female Taper

Lot or serial number

All lots before 2027/03/01

Model or catalog number

00-5850-046-20

00-5850-046-22

Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

UNITED STATES



H. Segmental System Variable Stiffness Stem Extension

Lot or serial number

All lots before 2027/03/01

Model or catalog number

00-5852-074-19

Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

UNITED STATES