Health product recall

ZILVER VASCULAR STENT WITH RAPID EXCHANGE (2016-02-09)

Starting date:
February 9, 2016
Posting date:
February 26, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57294

Affected Products

A. ZILVER VASCULAR STENT WITH RAPID EXCHANGE

Reason

Testing demonstrates a process used to manufacture the inner catheter to stylet joint of the delivery system doesn't meet the required specification on a consistent basis and there is a risk the joint may separate during deployment. A failure of this joint would represent a device malfunction and were it to occur during a procedure may result in a partial stent deployment. The clinical effect of a partial deployment may be a need for surgical intervention, or may result in vessel trauma or an embolic event

Affected products

A. ZILVER VASCULAR STENT WITH RAPID EXCHANGE

Lot or serial number

4020115

4265701

4470824

Model or catalog number

ZIVX5-18-125-9-40

ZIVX5-18-125-9-60

Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana

UNITED STATES