ZILVER VASCULAR STENT WITH RAPID EXCHANGE (2016-02-09)
- Starting date:
- February 9, 2016
- Posting date:
- February 26, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-57294
Affected Products
A. ZILVER VASCULAR STENT WITH RAPID EXCHANGE
Reason
Testing demonstrates a process used to manufacture the inner catheter to stylet joint of the delivery system doesn't meet the required specification on a consistent basis and there is a risk the joint may separate during deployment. A failure of this joint would represent a device malfunction and were it to occur during a procedure may result in a partial stent deployment. The clinical effect of a partial deployment may be a need for surgical intervention, or may result in vessel trauma or an embolic event
Affected products
A. ZILVER VASCULAR STENT WITH RAPID EXCHANGE
Lot or serial number
4020115
4265701
4470824
Model or catalog number
ZIVX5-18-125-9-40
ZIVX5-18-125-9-60
Companies
- Manufacturer
-
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES