Health product recall

Zeus IE Anesthesia Workstation (2014-05-30)

Starting date:
May 30, 2014
Posting date:
June 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40045

Recalled Products

Zeus IE Anesthesia Workstation

Reason

The device is designed to generate an audible and visual alarm when the measured value for spo2 falls below the alarm threshold set for the spo2 low alarm by the clinician. If the measured value continues to fall, the device is designed to continue to generate a spo2 low alarm. But, for devices with software version 1.03, if the sp02 measured value drops below the spo2 low alarm threshold by more than 10 percent, the audible and visual alarm for sp02 low will stop. If the value measured rises again, the alarm will be generated again as soon as the measured value is ten percent or less below the spo2-low alarm threshold. All displayed measurements of values and spo2 curves are performing as designed at any time.

Affected products

Zeus IE Anesthesia Workstation

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

MK04000

Companies
Manufacturer
Draeger Medical GMBH
Moislinger Allee 53-55,
Lubeck
23542
GERMANY