Zeus IE Anesthesia Workstation (2014-05-30)
- Starting date:
- May 30, 2014
- Posting date:
- June 18, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40045
Recalled Products
Zeus IE Anesthesia Workstation
Reason
The device is designed to generate an audible and visual alarm when the measured value for spo2 falls below the alarm threshold set for the spo2 low alarm by the clinician. If the measured value continues to fall, the device is designed to continue to generate a spo2 low alarm. But, for devices with software version 1.03, if the sp02 measured value drops below the spo2 low alarm threshold by more than 10 percent, the audible and visual alarm for sp02 low will stop. If the value measured rises again, the alarm will be generated again as soon as the measured value is ten percent or less below the spo2-low alarm threshold. All displayed measurements of values and spo2 curves are performing as designed at any time.
Affected products
Zeus IE Anesthesia Workstation
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
MK04000
Companies
- Manufacturer
-
Draeger Medical GMBH
Moislinger Allee 53-55,
Lubeck
23542
GERMANY