Xpert FLU/RSV XC (2016-04-11)

Starting date:
April 11, 2016
Posting date:
April 29, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58162

Affected Products

Xpert FLU/RSV XC

Reason

Seven customers reported cartridge performance with a higher number of invalid results than would typically be expected with the kit and its intended use specimen types. Cepheid investigated the complaints and found that the complaints were valid. The failure investigation identifies six kit lots that were affected. Only one of the lots were imported and sold in Canada.

Affected products

Xpert FLU/RSV XC

Lot or serial number

1000023147

Model or catalog number

GXFLU/RSV-10

Companies
Manufacturer

CEPHEID

904 CARIBBEAN DRIVE

SUNNYVALE

94089

California

UNITED STATES