This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

XPER FLEX CARDIO PHYSIOMONITORING SYSTEM-MAIN UNIT (2017-11-17)

Starting date:
November 17, 2017
Posting date:
December 8, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-65328

Affected products

XPER FLEX CARDIO PHYSIOMONITORING SYSTEM-MAIN UNIT

Reason

While in Full Disclosure (FD) playback, a user may inadvertently close the Full Disclosure Control Window using the ESC key function, instead of pressing the “X” icon in the upper right corner of this window. Use of the ESC key function will close the Full Disclosure Control Window, but Full Disclosure waveforms and numerics will remain on the display (i.e., the system will remain in Full Disclosure playback mode), preventing the display of real-time waveforms, numerics, and any visual alarms.

Affected products

XPER FLEX CARDIO PHYSIOMONITORING SYSTEM-MAIN UNIT

Lot or serial number

All lots

Model or catalog number

FC2020

Companies
Manufacturer
Invivo, a Div. of Philips Medical Systems
12151 Research Parkway
Orlando
32826
Florida
UNITED STATES