Health product recall

XN CHECK (2020-04-09)



Starting date:
April 9, 2020
Posting date:
May 1, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72931


Last updated:

Affected Products

XN CHECK

Reason

Sysmex America, Inc. (SAI) recognized a trend with complaints from XN-series customers reporting a decrease in the recovery of fluorescent platelet (PLT-F) and absolute Immature Platelet Fraction (IPF#) parameters with the XN CHECK, Level 1 hematology control Lot sequence 0062, 0063, 0064 and 0065. Lot 0062 and 0065 were distributed to Canadian customers beginning March 2, 2020.

Affected products

XN CHECK

Lot or serial number
  • 0062
  • 0065
Model or catalog number
  • 213499
Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN