Health product recall

XN CHECK (2018-01-04)

Starting date:
January 4, 2018
Posting date:
January 29, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65782

Affected products

XN CHECK

Reason

Sysmex has become aware of an increase in imprecision of lymphocyte (lym #, lym %) and Monocyte (Mono #, Mono %) counts with the XN Check hematology control. All other parameters on the assay sheet are not affected. Patient results are also unaffected.

Affected products

XN CHECK

Lot or serial number
  • 7317
  • 7319
Model or catalog number

213499

Companies
Manufacturer
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN