Health product recall

XN CHECK (2016-03-30)

Starting date:
March 30, 2016
Posting date:
April 15, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57934

Affected products

A. XN CHECK

Reason

Sysmex America Inc. has become aware that the IPF% level 2 expected range was incorrectly reported for XN check lot numbers 6067 and 6068. The IPF% level 2 mean values are correct. Applicable instruments are XN-series (XN-10, XN-20, XN-11, XN-21) analyzers. Assay sheets have been amended with the correct IPF% level 2 expected ranges. All other parameter means and expected ranges are correct. Use of the original assay sheet may falsely indicate that the control is recovering outside of the true expected range.

Affected products

A. XN CHECK

Lot or serial number

6067

6068

Model or catalog number

213499

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

JAPAN