This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

XN-10 and X-11 Modules (2017-04-04)

Starting date:
April 4, 2017
Posting date:
May 2, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63136

Affected Products

  1. XN-10 Module
  2. XN-11 Module

Reason

The protective plastic sheet that is intended to cover and protect electronic components inside the XN-series analyzers was not positioned correctly during manufacturing. Reagent from a leaking valve created a build-up of salt on an electronic component because of the misalignment of the protective sheet. The salt build-up created a salt bridge that allowed current to continue to flow, create heat, and eventually ignite adjacent resin parts inside the analyzer.

Affected products

A. XN-10 Module

Lot or serial number

11001 - 29656

29668 - 29786

29788-29798

29800-29801

29803

29804

29807

29812

29816-29867

Model or catalog number

XN-10

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN



B. XN-11 Module

Lot or serial number

A1001-A1124

A1127

A1130

A1135-A1140

A1142-A1144

A1146-A1149

A1151

A1153-A1159

A1161

A1163

A1171

Model or catalog number

XN-11

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN