This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

XIO RADIATION TREATMENT PLANNING SYSTEM (2015-08-28)

Starting date:
August 28, 2015
Posting date:
October 14, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55332

Affected Products

XIO RADIATION TREATMENT PLANNING SYSTEM

Reason

The manufacturer has discovered that XiO will calculate wedge dose in the region outside of the physical and the LINAC will deliver open dose. The patient will receive dose that is different from the planned dose and that dose difference can exceed 5%. There is a remote probability of serious injury resulting from this issue.

Affected products

XIO RADIATION TREATMENT PLANNING SYSTEM

Lot or serial number

Version 4.51 or higher

Model or catalog number

9200

Companies
Manufacturer

Elekta Business Area Software Systems Impac Medical Systems, Inc.

13723 RIVERPORT DRIVE
SUITE 100

Maryland Heights

63043