This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Xio Radiation Treatment Planning System (2013-12-20)

Starting date:
December 20, 2013
Posting date:
February 11, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37871

Recalled products

  1. Xio Radiation Treatment Planning System 

Reason

When the bolus is present, the effective depth (with bolus) to the weight point returned on the source data report is incorrect. The problem occurs when the bolus polygon incorrectly invades the patient contour and overlaps an area where the CT density is much different than bolus density. This is more likely to occur with convex patient surfaces.

Affected products

A. Xio Radiation Treatment Planning System 

Lot or serial number
  • XiO Release 4.50.00  
Model or catalog number
  • 9200
Companies
Manufacturer

Elekta Business Area Software Systems Impac Medical Systems Inc.

13723 Riverport Drive, Suite 100

Maryland Heights

63043

Missouri

UNITED STATES