XEVO TQ-S IVD Mass Spectrometer (2013-09-04)
- Starting date:
- September 9, 2013
- Posting date:
- September 18, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-35777
Recalled Product
XEVO TQ-S IVD Mass Spectrometer
Reason
The scroll backing pumps on the mass spectrometer are fitted with PTFE composite tip seals that wear down over time, generating a fine dust which is removed through routine maintenance. However, under certain conditions, excessive amounts of dust may be drawn into the analyser housing and may lead to a loss in sensitivity.
Affected products
XEVO TQ-S IVD Mass Spectrometer
Lot or serial number
- WAA429
- WAA455
- WAA471
- WAA714
Model or catalog number
- 186005453IVD