This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

XEN 45 Glaucoma Treatment System (2015-12-14)

Starting date:
December 14, 2015
Posting date:
January 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56556

Affected Products

A. XEN 45 Glaucoma Treatment System

Reason

Surgeons may encounter a higher than expected force to actuate slider mechanism on some XEN injectors when they are in the process of implanting the XEN 45 gel stent which may lead to ocular trauma especially when surgeon forces slider mechanism forward.

Affected products

A. XEN 45 Glaucoma Treatment System

Lot or serial number
  • 60815
  • 60869
Model or catalog number
  • 5507-001
Companies
Manufacturer

Aquesys Inc.

26970 Aliso Viejo Parkway, Suite 200

Aliso Viejo

92656

California

UNITED STATES