This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

VWR Pre-Calibrated Microhematocrit Tubes With Heparin (2014-08-11)

Starting date:
August 11, 2014
Posting date:
January 23, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43411

Recalled Products   

VWR Pre-Calibrated Microhematocrit Tubes With Heparin 

Reason

Drummond manufactures two products under the name of pre-cal. They are similar in appearance but have different OD's and ID's. In this instance the machine operator misread the specification and used the larger ID glass. Both of these tubes will deliver the same results when performing a hematocrit test. The pre-calibrated tube is used to determine a more consistent column height of blood at 60mm.

Affected products

VWR Pre-Calibrated Microhematocrit Tubes With Heparin 

Lot or serial number

1394

Model or catalog number

15401-560

Companies
Manufacturer

Drummond Scientific Company,

500 Parkway,

Broomall

19008

Pennsylvania

UNITED STATES