Volista Surgical Lights (2018-01-08)
- Starting date:
- January 8, 2018
- Posting date:
- January 29, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65780
Affected products
Volista Surgical Lights
Reason
There is a possibility that the central handle holder detaches from the cupola during use. We determined that the root cause of this malfunction is excessive tightening of the screws during mounting of the handle holder, resulting in the rivets loosening and detaching from the central handle holder ring. There is a remote chance that this issue may cause post-operative infection, foreign body granuloma or allergic reactions, if the particles fall into the patient's wound or on sterile equipment.
Affected products
Volista Surgical Lights
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- ARD568811903
- ARD568811912
- ARD568811913
- ARD568811952
- ARD568811961
- ARD568811963
- ARD568812912
- ARD568812913
- ARD568812952
- ARD568812963
- ARD568821960
- ARD568821961
Companies
- Manufacturer
-
Maquet S.A.
Avenue de la Pomme De Pin, Parc De Limere
Ardon, Orleans
45074
FRANCE