VITROS® XT Chemistry Product ALB-TP Slides
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
VITROS® XT Chemistry Product ALB-TP Slides |
1900148359 |
684 4293 |
Issue
QuidelOrtho™ has confirmed an issue involving VITROS® XT Chemistry Products ALB-TP slides which may create dust and debris within the microslide subsystem on VITROS® XT 3400 Chemistry Systems and VITROS® XT 7600 integrated systems. Dust and debris from XT ALB-TP slides may impact VITROS® Chemistry Products Na+ slides leading to a potential increase in non-reproducible, positively, or negatively biased Na+ results.
Recall start date: July 12, 2024
Additional information
Details
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive, Rochester, New York, United States, 14626
Get notified