VITROS® XT Chemistry Product ALB-TP Slides
Brand(s)
Last updated
Summary
Product
VITROS® XT Chemistry Product ALB-TP Slides
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
VITROS® XT Chemistry Product ALB-TP Slides | 1900148359 1900179815 1700120620 | 684 4293 |
Issue
QuidelOrtho™ has confirmed an issue involving VITROS® XT Chemistry Products ALB-TP slides which may create dust and debris within the microslide subsystem on VITROS® XT 3400 Chemistry Systems and VITROS® XT 7600 integrated systems. Dust and debris from XT ALB-TP slides may impact VITROS® Chemistry Products Na+ slides leading to a potential increase in non-reproducible, positively, or negatively biased Na+ results.
Recall start date: July 12, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive, Rochester, New York, United States, 14626
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75871
Get notified
Receive notifications for new and updated recalls and alerts by category.