Health product recall

Vitros White Correction Factor Slides (2014-01-10)

Starting date:
January 10, 2014
Posting date:
March 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38265

Recalled Products

Vitros White Correction Factor Slides 

Reason

OCD has initiated a recall as it has determined that one lot of Vitros White Correction Factor Slides was released with an incorrect assay value for the 460 nm wavelength filter. The labeling for the Vitros White Reference Slides lot 9052-0045-6358 incorrectly indicates an assay value of 0.3785 for the 460 nm filter versus the actual assay value of 0.8658. All other assay values listed are correct. The 460 nm wavelength is only used to determine the predicted concentration when processing Vitros TBIL slides or Vitros BUBC slides.

Affected products

Vitros White Correction Factor Slides 

Lot or serial number

9052-0045-6358

Model or catalog number

J02315

Companies
Manufacturer

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES