This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Vitros Systems (2014-04-11)

Starting date:
April 11, 2014
Posting date:
May 5, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39319

Recalled Products

A. Vitros Chemistry Products 5.1 FS Chemistry System - Analyzer
B. Vitros Chemistry Products 5.1 FS Chemistry System - Refurbished
C. Vitros Chemistry Products 5.1 FS Chemistry System - Instrument
D. Vitros 5600 Integrated System - Instrument
E. Vitros 5600 Integrated System
F.  Vitros 4600 Chemistry System - Hardware

Reason

OCD has issued a recall due to the potential for biased results when a multiple window (MW) code is associated with results generated by the Vitros 5600 Integrated system, Vitros 4600 Chemistry system and Vitros 5.1 Chemistry system. MW codes may be associated with results from any of the following Vitros Chemistry products using multiple point (Multipoint) rate measurements: Alt slides, ALKP slides, AST slides, CHE slides, CK slides, GGT slides, LDH slides. Multipoint rate assays measure enzyme activity via multiple readings which are used to generate a kinetic curve and calculate results. The MW algorithm assesses the validity of the kinetic curve. An MW code is generated if a single point exceeds predetermined limits, provided that all surrounding values are within acceptable limits.

Affected products

A. Vitros Chemistry Products 5.1 FS Chemistry System - Analyzer
 

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

6801375
 

Companies
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES

B. Vitros Chemistry Products 5.1 FS Chemistry System - Refurbished
 

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

6801890
 

Companies
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES

C. Vitros Chemistry Products 5.1 FS Chemistry System - Instrument
 

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

6801375
 

Companies
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES

D. Vitros 5600 Integrated System - Instrument
 

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

680 2413
 

Companies
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES

E. Vitros 5600 Integrated System
 

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

680 2413
 

Companies
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES

F.  Vitros 4600 Chemistry System - Hardware

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

680 2445

Companies
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES