This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

VITROS Immunodiagnostic Products Intact PTH (Reagent Pack & Calibrators)

Starting date:
October 6, 2016
Posting date:
November 3, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60780

Affected Products

Reason

Ortho confirmed that results obtained from VITROS Intact Parathyroid Hormone (iPTH) Reagent Packs are positively biased (i.e., falsely elevated) compared to an alternative commercially available method.

Ortho observed a positive bias of 40% for samples with iPTH concentrations

Ortho data demonstrated that this bias was consistent for all in-date lots.

Affected products

VITROS Immunodiagnostic Products Intact PTH (Reagent Pack & Calibrators)

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number
  • 680 2892
  • 680 2893
Companies
Manufacturer

Ortho-Clinical Diagnostics

Felindre Meadows, Pencoed, Bridgend, Great Britain

CF35 5PZ

UNITED KINGDOM


100pg>