VITROS Immunodiagnostic Products HBsAg Controls (2016-06-23)

Starting date:
June 23, 2016
Posting date:
July 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-59250

Affected products

A. VITROS Immunodiagnostic Products HBsAg Controls

Reason

Ortho Clinical Diagnostics (Ortho) has initiated this recall due to the potential for higher than expected signal/cutoff (s/c) results outside the control's expected range when using Vitros HBsAg controls (the negative control). The positive shift in results occurs prior to the 5 day opened-reconstituted stability product claim for Vitros HBsAg controls, lots 0860, 0870, 0880, 0890 and 0908. A positive shift with the positive control may also be observed, but the shift is within the control's expected range.

Affected products

A. VITROS Immunodiagnostic Products HBsAg Controls

Lot or serial number

0860

0870

0880

0890

0908

Model or catalog number

6800598