Health product recall

VITROS Immunodiagnostic Products Estradiol Reagent Pack (2016-07-15)

Starting date:
July 15, 2016
Posting date:
August 5, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-59704

Affected products

A. VITROS Immunodiagnostic Products Estradiol Reagent Pack

Reason

Ortho Clinical Diagnostics (Ortho) has initiated this recall due to the potential for medications that are derivatives of estrogen (e.g. fulvestrant, brand name: Faslodex) to interfere with estradiol immunoassays and cause positively biased sample results. Fulvestrant is an estrogen receptor antagonist medication that blocks estrogen from tumors. The drug is used in the treatment of patients with metastatic, postmenopausal, hormone receptor-positive breast cancer.

Affected products

A. VITROS Immunodiagnostic Products Estradiol Reagent Pack

Lot or serial number

1470

1490

1500

1510

1538

Model or catalog number

8552630

Companies
Manufacturer

ORTHO-CLINICAL DIAGNOSTICS

FELINDRE MEADOWS, PENCOED, BRIDGEND