Health product recall

VITROS Immunodiagnostic Products Anti-SARS-CoV-2 Total Reagent Pack (2020-07-22)

Starting date:
July 22, 2020
Posting date:
August 28, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73757

Last updated:
2020-08-28

Affected products

VITROS Immunodiagnostic Products Anti-SARS-CoV-2 Total Reagent Pack

Reason

A VITROS Anti-SARS-CoV-2 Total Reagent Pack in which the reagent bottles were swapped within the reagent pack. If the two reagent bottles are switched in their positions, the correct reaction scheme is not followed for processing the assay. This can result in:

  • False negative results for a sample that is reactive.
  • Calibration failures, if the affected pack is used during calibration. This could lead to a delay in testing.
  • Quality control failures for the reactive control, if the affected pack is used for testing quality control samples. This could lead to a delay in testing or releasing results.

Affected products

VITROS Immunodiagnostic Products Anti-SARS-CoV-2 Total Reagent Pack

Lot or serial number

0010 - 0037

Model or catalog number

619 9922

Companies
Manufacturer

Ortho-Clinical Diagnostics Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES