VITROS Immunodiagnostic Products Anti-SARS-CoV-2 Total Reagent Pack (2020-07-22)
- Starting date:
- July 22, 2020
- Posting date:
- August 28, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73757
Last updated:
2020-08-28
Affected products
VITROS Immunodiagnostic Products Anti-SARS-CoV-2 Total Reagent Pack
Reason
A VITROS Anti-SARS-CoV-2 Total Reagent Pack in which the reagent bottles were swapped within the reagent pack. If the two reagent bottles are switched in their positions, the correct reaction scheme is not followed for processing the assay. This can result in:
- False negative results for a sample that is reactive.
- Calibration failures, if the affected pack is used during calibration. This could lead to a delay in testing.
- Quality control failures for the reactive control, if the affected pack is used for testing quality control samples. This could lead to a delay in testing or releasing results.
Affected products
VITROS Immunodiagnostic Products Anti-SARS-CoV-2 Total Reagent Pack
Lot or serial number
0010 - 0037
Model or catalog number
619 9922
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES