Health product recall

VITROS Immunodiagnostic Products Anti-HBs Calibrators And Reagent

Last updated

Summary

Product
VITROS Immunodiagnostic Products Anti-HBs Calibrators And Reagent
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

VITROS Immunodiagnostic Products Anti-HBs Calibrators And Reagent

5111
5120

152 4693
178 7753

Issue

QuidelOrtho's internal stability testing of VITROS anti-HBs calibrators detected a gradual reduction in calibration signal for calibrator levels 2 and 3 over the shelf life of the product for the specified lots. This reduction in signal may affect the shape of the calibration curve generated during calibration, compared to the calibration curve generated earlier in the product's shelf life. A change in the shape of the calibration curve may cause patient and QC samples tested using the associated VITROS® immunodiagnostic products anti-HBs reagent pack to report results as falsely elevated, potentially causing negative results to report as borderline, and borderline results to report as positive. This issue is detectable by performing QC as results generated will be higher than expected and/or outside the expected limits for QC, depending on the age of the lot and amount of signal reduction experienced in calibrator levels 2 and 3.

Recall start date: May 21, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Immunology
Companies
Ortho-Clinical Diagnostics
Felindre Meadows, Bridgend, Great Britain, CF35 5PZ
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75591

Get notified