Vitros Immunodiagnostic, Integrated and Chemistry System (2014-11-20)
- Starting date:
- November 20, 2014
- Posting date:
- December 11, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Hospitals
- Identification number:
- RA-42845
Recalled Products
A) Vitros 3600 Immunodiagnostic System (Licence # 78572)
B) Vitros 5600 Integrated System (Licence # 78571)
C) Vitros 5600 Integrated System (Licence # 80701)
D) Vitros 3600 Immunodiagnostic System (Licence # 80702)
E) Vitros 4600 Chemistry System
Reason
Ortho Clinical Diagnostics (OCD) has identified an anomaly with Vitros system software version 3.1 and below. OCD determined that the software may not properly identify an expired calibration. Calibration status is monitored by the system software which is based upon the calibration interval and the date of the last calibration. OCD's investigation confirmed that if the anomaly occurs, a calibration may expire, but the system may not alert the operator and not flag results with a CE sample result code (i.e. Calibration expired). The system may unexpectedly generate results even though it is configured to not report results when a calibration has expired.
Affected products
A) Vitros 3600 Immunodiagnostic System (Licence # 78572)
Lot or serial number
36000130
36000264
36000627
Model or catalog number
680 2783
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
B) Vitros 5600 Integrated System (Licence # 78571)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
680 2413
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
C) Vitros 5600 Integrated System (Licence # 80701)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
680 2413
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
D) Vitros 3600 Immunodiagnostic System (Licence # 80702)
Lot or serial number
36000130
36000264
36000627
Model or catalog number
680 2783
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
E) Vitros 4600 Chemistry System
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
680 2445
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES