Health product recall

Vitros Hand-held Barcode Scanner (accessory to Vitros 3600 Immunodiagnostic System, Vitros 5600 Integrated System, Vitros 5,1 & 4600 Chemistry System) (2014-12-30)

Starting date:
December 30, 2014
Posting date:
January 14, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43287

Recalled Products

Vitros Hand-held Barcode Scanner (accessory to Vitros 3600 Immunodiagnostic System, Vitros 5600 Integrated System, Vitros 5,1 & 4600 Chemistry System)

Reason

Ortho Clinical Diagnostics (OCD) has received a report that when using a hand-held barcode scanner in combination with use of the batch programming option for programming samples, the scanner unexpectedly skips sample cup positions. For example, a sample tube barcode is scanned for the cup 1 position, the cursor in the batch programming screen skips the cup 2 position and advances to the cup 3 position. If this anomaly occurs, there is a potential for mis-associated results.

Affected products

Vitros Hand-held Barcode Scanner (accessory to Vitros 3600 Immunodiagnostic System, Vitros 5600 Integrated System, Vitros 5,1 & 4600 Chemistry System)

Lot or serial number

0914F11984

0914F11985

Model or catalog number

6844210

Companies
Manufacturer

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES