Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61206

Reason

Ortho's investigation confirmed that samples diluted (3x recommended dilution or higher) using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3 or a low c-reactive protein (CRP) concentration sample had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result.

Affected products

Vitros Chemistry Products System-CRP Slides

Lot or serial number

More than 10 lots, contact manufacturer.

Model or catalog number

  • 192 6740
  • 809 7990

Companies

Manufacturer

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester

14626

New York