Health product recall

Vitros Chemistry Products Na+ Slides

Starting date:
December 21, 2016
Posting date:
February 1, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62044

Reason

Ortho Clinical Diagnostics (Ortho) identified the potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ cartridges warmed between 1.5 to 8 hours then placed on the analyzer. Ortho has determined that cartridges require a minimum of 8 hours of warm up at room temperature to help ensure the slides have acceptable performance throughout the 10 day on-analyzer storage limit.

Affected products

Vitros Chemistry Products Na+ Slides

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

837 9034

Companies

Manufacturer

Ortho-Clinical Diagnostics Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES