This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Vitros Chemistry Products HPT Reagent

Starting date:
November 2, 2012
Posting date:
November 26, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15894

Recalled Products

  1. Vitros Chemistry Products HPT Reagent

Reason

Ortho Clinical Diagnostic (OCD) has received complaints of the inability to calibrate the Vitros Chemistry products HPT reagent. OCD has also received reports of an increase in U91-274 condition codes. Although all affected lots met all release testing specifications prior to distribution, an OCD investigation has confirmed an increased occurrence of condition codes and calibration failures when using these lots over a period of time. OCD testing has found that as the reagent ages, a potential for biased results exists if you are able to obtain a valid calibration.

Affected products

A. Vitros Chemistry Products HPT Reagent

Lot or serial number

1533-20-2314, 1533-20-2314

Model or catalog number

680 2214

Companies
Manufacturer

Ortho-Clinical Diagnostics Inc.