Health product recall

Vitros Chemistry Products COCM Reagent (2013-10-04)

Starting date:
October 4, 2013
Posting date:
November 25, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36847

Recalled products

A. Vitros Chemistry Products COCM Reagent 

Reason

OCD has received customer complaints of a downward drift/shift in quality control results or low quality control results when using Vitros chemistry products COCM reagent. This drift/shift will likely be observed within approximately 3-14 days of calibration. Patient results may be affected in the same manner. Our investigation has shown that recalibration will be initially effective in resolving the issue, but that increased calibration frequency may no longer be effective over time.

OCD internal testing has confirmed that this issue occurs will all lots of Vitros COCM reagent that are within expiration dating. As a result, customers are being asked to discontinue using and discard all Vitros COCM reagent remaining in their inventory.

Affected products

A. Vitros Chemistry Products COCM Reagent 

Lot or serial number
  • 1525-09-2754
  • 1525-10-2901
  • 1525-10-3062
Model or catalog number
  • 680 1995
Companies
Manufacturer

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES