Vitros Chemistry Products COCM Reagent (2013-10-04)
- Starting date:
- October 4, 2013
- Posting date:
- November 25, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-36847
Recalled products
A. Vitros Chemistry Products COCM Reagent
Reason
OCD has received customer complaints of a downward drift/shift in quality control results or low quality control results when using Vitros chemistry products COCM reagent. This drift/shift will likely be observed within approximately 3-14 days of calibration. Patient results may be affected in the same manner. Our investigation has shown that recalibration will be initially effective in resolving the issue, but that increased calibration frequency may no longer be effective over time.
OCD internal testing has confirmed that this issue occurs will all lots of Vitros COCM reagent that are within expiration dating. As a result, customers are being asked to discontinue using and discard all Vitros COCM reagent remaining in their inventory.
Affected products
A. Vitros Chemistry Products COCM Reagent
Lot or serial number
- 1525-09-2754
- 1525-10-2901
- 1525-10-3062
Model or catalog number
- 680 1995
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES