VITROS Chemistry Products Calcium (Ca) Slides (2016-05-19)
- Starting date:
- May 19, 2016
- Posting date:
- June 16, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Hospitals, Healthcare Professionals
- Identification number:
- RA-58896
Affected products
- VITROS Chemistry Products Calcium (Ca) Slides
Reason
Ortho Clinical Diagnostics (Ortho) has initiated this recall due to a potential for negatively biased serum sample results to be generated when using VITROS Ca Slides, Gen 29.
Ortho confirmed an average negative bias of -0.45 mg/dl (-0.11 mmol/l) within the serum reference interval (8.4 - 10.2 mg/dl or (2.10 - 2.55 mmol/l) when using VITROS Ca Slides, Gen 29.
Plasma and serum samples are similarly affected as both sample types use the same calibration parameters (math) when results are produced. Though unconfirmed, to date, Ortho has no indication that urine sample results, generated when using VITROS Ca Slides, Gen 29, are affected. As a result, customers are being instructed to discontinue the use of VITROS Ca Slides, Gen 29.
Affected products
A. VITROS Chemistry Products Calcium (Ca) Slides
Lot or serial number
- 0329-0528-6999
- 0329-0528-8902
- 0329-0528-9167
- 0329-0528-9176
- 0329-0528-9265
Model or catalog number
- 1450261
Companies
- Manufacturer
-
Ortho Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES