This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Vitros Chemistry Products (2013-10-18)

Starting date:
October 18, 2013
Posting date:
November 29, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36879

Recalled Products

A. Vitros Chemistry Products Sodium Slides (NA+)

B. Vitros Chemistry Products Calibrator Kit 2

Reason

OCD has confirmed that positively biased serum or plasma sample results may occur when using Vitros NA+ slides, Gens 7-13. Our data determined that the largest bias was observed when using slides from Gens 7-13 when calibrated with Vitros calibrator kit 2, lots 212 and 292. The amount of bias is dependent upon multiple sources of variability that include but are not limited to the slide gen, calibrator kit lot, calibration event and Vitros system. It is important to note that not all combinations of Vitros NA+ slide Gens and calibrator kit lots resulted in biased NA+ results.

Customers may continue to use their remaining inventory of Gens 7-13 providing the following criteria are met:/p>

  1. calibration is successful
  2. quality control results are within the established range of means/acceptable limits.
  3. distribution of normal results for serum or plasma samples are centered within the established reference interval for their laboratory.

Affected products

A. Vitros Chemistry Products Sodium Slides (NA+)

Lot or serial number

GENs 7 - 13

Model or catalog number

8379034

Companies
Manufacturer

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES



B.     Vitros Chemistry Products Calibrator Kit 2

Lot or serial number

212

292

Model or catalog number

1662659

Companies
Manufacturer

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES