Vitros Chemistry Products (2013-10-18)
- Starting date:
- October 18, 2013
- Posting date:
- November 29, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-36879
Recalled Products
A. Vitros Chemistry Products Sodium Slides (NA+)
B. Vitros Chemistry Products Calibrator Kit 2
Reason
OCD has confirmed that positively biased serum or plasma sample results may occur when using Vitros NA+ slides, Gens 7-13. Our data determined that the largest bias was observed when using slides from Gens 7-13 when calibrated with Vitros calibrator kit 2, lots 212 and 292. The amount of bias is dependent upon multiple sources of variability that include but are not limited to the slide gen, calibrator kit lot, calibration event and Vitros system. It is important to note that not all combinations of Vitros NA+ slide Gens and calibrator kit lots resulted in biased NA+ results.
Customers may continue to use their remaining inventory of Gens 7-13 providing the following criteria are met:/p>
- calibration is successful
- quality control results are within the established range of means/acceptable limits.
- distribution of normal results for serum or plasma samples are centered within the established reference interval for their laboratory.
Affected products
A. Vitros Chemistry Products Sodium Slides (NA+)
Lot or serial number
GENs 7 - 13
Model or catalog number
8379034
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
B. Vitros Chemistry Products Calibrator Kit 2
Lot or serial number
212
292
Model or catalog number
1662659
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES