VITROS Chemistry Kits (2015-09-29)
- Starting date:
- September 29, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-55638
Affected Products
A. VITROS Chemistry Products Calibrator Kit 9
B. VITROS Chemistry Products Calibrator Kit 9
Reason
Ortho Clinical Diagnostics, Inc. (OCD) initiated this recall due to the potential for an increase in U90-382 or 6LU condition codes (wash errors) when using Vitros Calibrator Kit 9, Lot 0954. When a U90-382 or 6LU condition code occurs, the Vitros System suppresses the result and no result is reported. The increase in condition codes primarily occurs when using Vitros DGXN and PHYT slides. All numerical results obtained are valid providing that no wash errors occurred and quality control results were within acceptable limits.
Affected products
A. VITROS Chemistry Products Calibrator Kit 9
Lot or serial number
0954
Model or catalog number
8568040
Companies
- Manufacturer
-
Ortho Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
B. VITROS Chemistry Products Calibrator Kit 9
Lot or serial number
954
Model or catalog number
8568040
Companies
- Manufacturer
-
Ortho Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES