VITROS 4600 and 5,1 FS Chemistry Systems (2015-11-12)
- Starting date:
- November 12, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56286
Affected Products
- VITROS 5,1 FS Chemistry System
- VITROS 5,1 FS Chemistry System
- VITROS 4600 Chemistry System
Reason
Ortho Clinical Diagnostics (OCD) has issued this recall due to a software anomaly that may potentially occur on Vitros 4600 or 5,1 FS Systems. OCD recently identified that under very specific conditions, the temperature in the Microwell Subsystem (included on Vitros 3600 and 5600 Systems) may potentially be out of range without the system properly alerting the operator. Although Vitros 4600 and 5,1 FS systems do not include a Microwell Subsystem, it utilizes the same software as Vitros 3600 and 5600 Systems. OCD has estimated that there is a less than 1 in 14 million possibility that this anomaly could occur in the Microslide or Microtip subsystems included on Vitros 4600 or 5,1 FS Systems.
Affected products
A. VITROS 5,1 FS Chemistry System
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
680 1375
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
B. VITROS 5,1 FS Chemistry System
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
680 1375
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
C. VITROS 4600 Chemistry System
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
680 2445
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES