Health product recall

VITROS 4600 and 5,1 FS Chemistry Systems (2015-11-12)

Starting date:
November 12, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56286

Affected Products

  1. VITROS 5,1 FS Chemistry System
  2. VITROS 5,1 FS Chemistry System
  3. VITROS 4600 Chemistry System

Reason

Ortho Clinical Diagnostics (OCD) has issued this recall due to a software anomaly that may potentially occur on Vitros 4600 or 5,1 FS Systems. OCD recently identified that under very specific conditions, the temperature in the Microwell Subsystem (included on Vitros 3600 and 5600 Systems) may potentially be out of range without the system properly alerting the operator. Although Vitros 4600 and 5,1 FS systems do not include a Microwell Subsystem, it utilizes the same software as Vitros 3600 and 5600 Systems. OCD has estimated that there is a less than 1 in 14 million possibility that this anomaly could occur in the Microslide or Microtip subsystems included on Vitros 4600 or 5,1 FS Systems.

Affected products

A. VITROS 5,1 FS Chemistry System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

680 1375

Companies
Manufacturer

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES



B. VITROS 5,1 FS Chemistry System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

680 1375

Companies
Manufacturer

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES



C. VITROS 4600 Chemistry System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

680 2445

Companies
Manufacturer

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES