This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Vitek MS-CHCA Matrix and Vitek MS-DS Target Slides (2014-02-27)

Starting date:
February 27, 2014
Posting date:
April 4, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38663

Recalled products

  1. Vitek MS-CHCA Matrix
  2. Vitek MS-DS Target Slides

Reason

It has been confirmed that one of the positive QC strains, Enterococcus faecalis ATCC 19433, was not reproducible when tested on Vitek MS following culture on the Trypcase Soy Agar + 5% sheep blood (TSS) media. During development, this Enterococcus faecalis ATCC 19433 strain was tested after culture on the Colombia Agar + 5% sheep blood media. Therefore, the culture of this strain on the Trypcase Soy Agar + 5% sheep blood (TSS) media was never validated.

Affected products

A. Vitek MS-CHCA Matrix

Lot or serial number

Not applicable

Model or catalog number
  • 411071
Companies
Manufacturer

Biomerieux SA

chemin de l'Orme

Marcy l'Etoile

69280

FRANCE



B. Vitek MS-DS Target Slides

Lot or serial number

Not applicable

Model or catalog number
  • 410893
Companies
Manufacturer

Biomerieux SA

chemin de l'Orme

Marcy l'Etoile

69280

FRANCE