Health product recall

VITEK 2 Test Cards (2019-01-14)

Starting date:
January 14, 2019
Posting date:
February 15, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69030

Affected Products

  • A. VITEK 2 Compact 30 System - susceptibility test cards
  • B. VITEK 2 Compact 60 System - susceptibility test cards
  • C. VITEK 2 System - susceptibility test cards
  • D. VITEK 2 ANC ID
  • E. VITEK 2 GN Card

Reason

A potential issue was identified related to the top seal of card pouches containing the VITEK 2 test cards. BioMérieux has determined that the integrity of some of the VITEK 2 test card top seal may be compromised.  A compromised test card top seal can impact card reagents due to the entry of moisture. BioMérieux investigation concluded that the pouch top seal of the card lots manufactured on a specific card poucher (poucher 1) between 10 Aug. 2018 and 29 Aug. 2018 is potentially compromised.

Affected products

A. VITEK 2 Compact 30 System - susceptibility test cards

Lot or serial number

6260813203

Model or catalog number

413 143

Companies
Manufacturer

Biomerieux Inc.

100 Rudolphe Street

Durham, North Carolina

27712

UNITED STATES



B.  VITEK 2 Compact 60 System - susceptibility test cards

Lot or serial number

2750809103

2750823203

Model or catalog number

415 670

Companies
Manufacturer

Biomerieux Inc.

100 Rudolphe Street

Durham, North Carolina

27712

UNITED STATES



C. VITEK 2 System - susceptibility test cards

Lot or serial number

2880809203

5420809203

5420812203

 

Model or catalog number

420 739

421 040

 

Companies
Manufacturer

Biomerieux Inc.

100 Rudolphe Street

Durham, North Carolina

27712

UNITED STATES



D. VITEK 2 ANC ID

Lot or serial number

2440811203

Model or catalog number

21347

Companies
Manufacturer

Biomerieux Inc.

100 Rudolphe Street

Durham, North Carolina

27712

UNITED STATES



E. VITEK 2 GN Card

Lot or serial number

2410602403

2410632203

2410636203

Model or catalog number

21341

Companies
Manufacturer

Biomerieux Inc.

100 Rudolphe Street

Durham, North Carolina

27712

UNITED STATES