This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

VITEK 2 COMPACT 30 SYSTEM - SUSCEPTIBILITY TEST C (2015-03-30)

Starting date:
March 30, 2015
Posting date:
April 30, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53165

Recalled Products

A. VITEK 2 COMPACT 30 SYSTEM - SUSCEPTIBILITY TEST C

Reason

Out of range low (OORL) flucytosine (FCT) QC results for Candida krusei ATCC 6258, the package insert QC strain, has been identified in association with the Vitec 2 AST-YS06 and AST-YS07 test kits.

Affected products

A. VITEK 2 COMPACT 30 SYSTEM - SUSCEPTIBILITY TEST C

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

412 610

414 967

Companies
Manufacturer

Biomerieux Inc.

595 Anglum Road

Hazelwood

Missouri

UNITED STATES