Health product recall

VITEK 2 Compact 30 System - Instrument (2017-03-02)

Starting date:
March 2, 2017
Posting date:
April 3, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66306

Affected products

VITEK 2 COMPACT 30 SYSTEM - INSTRUMENT

Reason

Several complaints were received for invalid calibration with low calibrator s1 while using VIDAS LH. Furthermore, tests performed internally confirmed a decrease of the strips signal overtime for the lot 1006041540. The defect was highlighted with complaints for invalid calibration, but concerns all tests performed (calibration, and patient's samples testing). The investigation concluded that this could have an impact on patient's results.

Affected products

VITEK 2 COMPACT 30 SYSTEM - INSTRUMENT

Lot or serial number

1006041540

Model or catalog number

30406

Companies
Manufacturer
BioMérieux S.A
376 Chemin De L'orme
Marcy-l'etoile
69280
FRANCE