This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Vital-Port Vascular Access Ports and Systems (2017-11-28)

Starting date:
November 28, 2017
Posting date:
December 21, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-65458

Affected products

  1. VITAL-PORT VASCULAR ACCESS PORT
  2. VITAL-PORT MRI VASCULAR ACCESS SYSTEM PORT
  3. VITAL-PORT MINI VASCULAR ACCESS
  4. VITAL-PORT PETITE VASCULAR ACCESS PORT
  5. VITAL-PORT MRI PETITE VASCULAR ACCESS
  6. VITAL-PORT PETITE DUAL LUMEN VASCULAR ACCESS

Reason

The non-coring needle included with the sets has the potential to dislodge silicone from the septum of the port component. Hypothetically, this could lead to embolization of a fragment or leakage of medication/injectable fluids into surrounding tissues. Neither of these potential consequences has been reported.

Affected products

A. VITAL-PORT VASCULAR ACCESS PORT

 

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

IP-7110

IP-S6010

IP-S6110

IP-S7010

IP-S7110

IP-S9010

IP-S9110

Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana

UNITED STATES



B. VITAL-PORT MRI VASCULAR ACCESS SYSTEM PORT

 

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

IP-7112

IP-9112

IP-S6112

IP-S7112

IP-S9012

IP-S9112

Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana

UNITED STATES



C. VITAL-PORT MINI VASCULAR ACCESS

 

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

IP-5116

IP-S5116

IP-S5116W

Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana

UNITED STATES



D. VITAL-PORT PETITE VASCULAR ACCESS PORT

 

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

IP-6113

IP-S6013

IP-S6113

IP-S6113-MPIS-NT

Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana

UNITED STATES



E. VITAL-PORT MRI PETITE VASCULAR ACCESS

 

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

IP-S5118

IP-S6118

IP-S6118-MPIS-NT

Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana

UNITED STATES



F. VITAL-PORT PETITE DUAL LUMEN VASCULAR ACCESS

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

IP-S7029

IP-S7129

Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana

UNITED STATES