Video Telescope
- Starting date:
- February 21, 2017
- Posting date:
- March 24, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-62752
Reason
Olympus has initiated this recall after discovering that the ENDOEYE distal end can become abnormally hot if the temperature sensor at the device's tip is damaged. Although no customer complaints have been received referring to this issue and thus manufacturer is not aware of any patient or user injury, excessive heating of the ENDOEYE distal end could result in patient or user injury. To prevent a potential risk to patient or user health, Olympus is undertaking this action to repair the devices by disabling a specific feature on the video telescopes ("fog-free function") to prevent excessive heating of the ENDOEYE distal end under certain conditions.
Affected products
Video Telescope
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
WA50040A
WA50042A
Companies
- Manufacturer
-
Olympus Winter & IBE GMBH
Kuehnstrasse 61
Hamburg
22045
GERMANY