VIDAS Protein C & TOXO IGG Avidity

Brand(s)
Last updated

Summary

Product
VIDAS Protein C & TOXO IGG Avidity
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or serial number

Model or catalog number

VIDAS - PROTEIN C ASSAY

1008571770
1008662180
1008801740

30115

VIDAS TOXO IGG AVIDITY

1008721750

30 222

Issue

Based on an unusual rate of complaints from the field for "Substrate error" leading to a potential delayed results on different  References tested on all  Systems, Biomerieux initiated an investigation. The substrate error issue was confirmed on all lots of  Immuno-Assays manufactured with substrate batches using the common lot of raw material (4-MUP). The only potential risk associated with the substrate error is a delayed result as you need to run another test. There is no risk of false results. The investigation demonstrates that the issue occurs over the product shelf-life. The analyzes of the kinetic model allowed to define a revised expiry date for each impacted lot of  Immuno-Assays finished products. When used until the revised expiry date, the product continues to perform per its registered performance specifications. Therefore, corrective actions involving a revised expiration date for all lots of clinical  Immunoassay products are required to ensure that the specified products will continue to perform per their labelled performance specifications.

Recall date: 2021-10-19

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies

BIOMERIEUX SA,

376, CHEMIN DE L'ORME, MARCY-L'ETOILE

FRANCE

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type III
Identification number
RA-63556