This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

VIDAS - ESTRADIOL II ASSAY (2016-07-12)

Starting date:
July 12, 2016
Posting date:
August 5, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59696

Affected products

A. VIDAS - ESTRADIOL II ASSAY

Reason

The drug fulvestrant cross react with antibodies used in immunoassays on the market. This is due to the fact that the fulvestrant molecule has a similar chemical structure to estradiol which leads to an overestimation of estradiol level.

Affected products

A. VIDAS - ESTRADIOL II ASSAY

Lot or serial number

1004343690

1004403350

1004666000

Model or catalog number

30431

Companies
Manufacturer

Biomerieux Sa

Chemin De L'orme

Marcy-l'etoile

69280

FRANCE