This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Vertex Epic Camera / Pegasys Ultra Imaging (2014-03-07)

Starting date:
March 7, 2014
Posting date:
April 4, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38803

Recalled Products

Vertex Epic Camera / Pegasys Ultra Imaging

Reason

Philips Healthcare received a report from the field about an Adac Vertex Plus system. During clinical use in the relative 180 degree configuration of the detector heads, the radius drive belt in the superior positioned head slipped off the idler pulley, allowing the detector head to drift down slowly (approximately 1 cm per minute) towards the patient. This resulted in the operator having to perform an emergency removal or the patient from the system.

Affected products

Vertex Epic Camera / Pegasys Ultra Imaging

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

AMT-VERTEXE
AMT-VERTEXP

Companies
Manufacturer
Philips Medical Systems (Cleveland) Inc.
3860 NORTH FIRST STREET,
SAN JOSE, CALIFORNIA
95134
UNITED STATES