Health product recall

Veritas Advanced Infusion Pack

Last updated

Summary

Product
Veritas Advanced Infusion Pack
Issue
Medical devices - Performance
What to do

Contact manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Veritas Advanced Infusion Pack

60330206

VRT-AI

Issue

Johnson & Johnson vision has initiated this action due to the potential for the irrigation luer to crack or break. A crack or break of the irrigation luer could lead to reduced irrigation pressure during surgery, associated with an unstable anterior chamber. This hazard could result in patient harm such as, but not limited to, capsule tear. Additionally, there could be a delay in surgery if a leakage is noticed during or prior to surgery and the pack must be switched out. As of August 29, 2022, there have been forty-one (41) product complaints, including one (1) adverse event, associated with this issue.

Recall start date: September 7, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Ophthalmology
Companies

Johnson & Johnson Surgical Vision, Inc.

1700 East St. Andrew Place, Santa Ana, California, United States, 92705

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72017

Get notified

Receive emails about new and updated recall and safety alerts.