(VEPTR) VERTICAL EXPANDABLE PROSTHETIC TITANIUM RIB CAUDAL RIB SUPPORTS (2015-08-04)
- Starting date:
- August 4, 2015
- Posting date:
- August 24, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54774
Affected Products
A. (VEPTR) VERTICAL EXPANDABLE PROSTHETIC TITANIUM RIB CAUDAL RIB SUPPORTS
Reason
This recall is being initiated because the last hole on the VEPTR Caudal Rib Supports, subject to this recall, may be partially formed. Fully lengthening the construct to the partially formed last hole may result in mechanical failure of the construct due to extension bar breakage, which could require unplanned surgical intervention to exchange components. In addition to the risks inherent to re-operation, potential effects on the patient could also include pain, loss of deformity correction, and/or soft tissue/visceral injury.
Affected products
A. (VEPTR) VERTICAL EXPANDABLE PROSTHETIC TITANIUM RIB CAUDAL RIB SUPPORTS
Lot or serial number
>100 NUMBERS, CONTACT MFG
Model or catalog number
497.087
497.088
497.089
497.093
497.094
497.095
497.098
497.099
497.100
497.225
497.226
497.227
497.228
497.229
497.230
497.231
497.232
497.233
497.234
497.235
497.236
497.237
497.238
497.239
497.241
497.242
497.243
497.244
497.245
497.246
497.247
497.248
497.249
Companies
- Manufacturer
-
Synthes USA LLC
1101 SYNTHES AVENUE
Monument
80132
Colorado
UNITED STATES