VENTANA® anti-Cd10 (SP67) Rabbit Monoclonal Primary Antibody
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
VENTANA® anti-Cd10 (SP67) Rabbit Monoclonal Primary Antibody |
K14266 |
790-4506 |
Issue
Roche has received complaints regarding high background and off-target staining with certain lots of the VENTANA® anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody resulting in a risk of false positive results.
Subsequent investigation identified that this high background and off-target staining can interfere with slide interpretation when using the recommended method sheet OptiView DAB IHC Detection Kit (product no.760-700 / 06396500001) / OptiView Amplification Kit (product no. 760-099 / 06396518001) and the ultraView Universal DAB Detection Kit (product no.760-500 i 05269806001) /Amplification Kit (product no. 760-080 / 05266114001) protocol.
Recall start date: July 11, 2024
Additional information
Details
Ventana Medical Systems Inc
1910 E Innovation Park Drive, Tucson, Arizona, United States, 85755
Get notified
Receive emails about new and updated recall and safety alerts.