Displaying 1 - 4 of 4 items.
During internal comparison studies, unacceptable light staining with some VENTANA PD-L1 (SP142) on-markets lots were confirmed which affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and…
RecallHealth product recall | 2022-12-14
This recall is regarding fluid leaks within the middle stainer module of the Ventana he 600 that likely resulted in electrical shorts at the plate heater
connections that then escalated into fires.
Recall start date: June 16, 2022
RecallHealth product recall | 2022-07-18
Internal stability testing was performed with the Pas Staining Kit where it was confirmed that the stability of 24 months could not be met. The expiration date of the product was reduced to 12 months after the manufacturing date.
Recall start date:…
RecallHealth product recall | 2022-07-18
The manufacturer reported a fluid leak within the Benchmark Ultra instruments that created an electrical short on the instrument's general purpose input/output (GPIO) board's printed circuit board (PCB). The event was localized to the J1 connector of the…
RecallHealth product recall | 2022-02-08