Health product recall

Ventana HE 600 System

Last updated

Summary

Product
Ventana HE 600 System
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalogue number

Ventana He 600 System

More than 10 numbers, contact manufacturer.

06917259001

Issue

This recall is regarding fluid leaks within the middle stainer module of the Ventana he 600 that likely resulted in electrical shorts at the plate heater
connections that then escalated into fires.

Recall start date: June 16, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies

Ventana Medical Systems Inc

1910 E Innovation Park Drive, Tucson, Arizona, United States, 85755

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-64372

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe