Venous Hardshell Cardiotomy Reservoir (2019-09-12)
- Starting date:
- September 12, 2019
- Posting date:
- September 20, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71027
Last updated:
Affected Products
Venous Hardshell Cardiotomy Reservoir
Reason
During verification testing of Adult/Small Adult VKMOs and VHK a potential impairment of the sterile packaging pouches of the accessories was detected, for products manufactured after March 4th 2019. Under unfavorable transport conditions, excessive movement of the device and its accessories in the carton can lead to stress points that could compromise the sterile barrier of the packaging pouches.
Affected products
Venous Hardshell Cardiotomy Reservoir
Lot or serial number
92278863
Model or catalog number
VHK 71000
Companies
- Manufacturer
-
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY